The EGA30 series is a versatile CO2 and humidity monitor designed for various applications such as greenhouse monitoring, indoor air quality, and process control. It features a wall-mounted CO2 infra-red gas analyser with advanced sensors for water vapour measurement.

Key Features
  • Accurate CO2 infra-red gas analyser with a range of 0-2000ppm and 1ppm resolution.
  • H2O measurement using laser-trimmed sensors with a range of 0-100% RH and 1% RH resolution.
  • Integral sample pump with a flow rate of at least 500cc/min.
  • Auto logging to a 2Gb SD memory card, providing extensive data storage capacity.
  • Two user-configurable alarm relays and seven analogue inputs for external sensors.
  • USB and analogue output for data download and integration.

Installation and Operation
The EGA30 is designed for quick and easy installation, taking approximately 15 minutes. It includes a two-part bracket for wall mounting and prewired electrical connections. The device features a thermostatically controlled infra-red gas analyser for stable operation across various temperatures, preventing condensation in the analysis cell.

Durability and Flexibility
The EGA30 is robust and durable, with a protective hinged front panel to prevent dirt, dust, and water ingress. It also features a long-life sample pump and an integral soda lime column for auto-zeroing the gas analyser, ensuring long-term calibration stability.

Data Management
Data is stored in an ASCII CSV format and can be downloaded via USB or by removing the SD card. The device supports flexible data logging, initiated by various triggers such as analogue input levels or relay actuation.

Technical Specifications
  • CO2 measurement range: 0-2000ppm, resolution: 1ppm.
  • H2O measurement range: 0-100% RH, resolution: 1% RH.
  • Atmospheric pressure range: 600-1100 mbar, accuracy: -4/+2 mbar.
  • Dimensions: Width 287mm x Depth 191mm x Height 168mm.
  • IP rating: 42.

Applications
  • Greenhouse monitoring and control.
  • Growth chamber monitoring and control.
  • Indoor air quality monitoring.
  • Process control.
  • Atmospheric air monitoring.

Company Information
ADC BioScientific Ltd is a leader in gas analysis technology, providing high-quality non-dispersive infra-red gas analysers for over 25 years. The company is committed to continuous improvement and innovation in its product offerings.
See more

Catalog excerpts

EGA30 Wall Mounted Analyser-1

EGA30 series A versatile CO2 and humidity monitor The EGA30 is a wall mounted CO2 infra-red gas analyser with advanced sensors for water vapour measurement. 2 user configurable alarm relays The EGA30 features a thermostatically controlled infrared gas analyser, ensuring stability over a wide range of operating temperatures. By controlling the temperature above ambient levels we remove the possibility of humid sample gasses condensing inside the gold plated analysis cell. H2O measurement is by the latest technology, laser trimmed sensors. All readings are automatically compensated for changes in atmospheric pressure. Auto logging to SD memory card Assured accuracy Integral sample pump For over 25 years ADC’s name has been synonymous with quality gas analysis. In thousands of laboratories, all over the world, researchers are using ADC technology to accurately measure concentrations of analytical gasses. H2O by laser trimmed sensor ADC BioScientific Ltd: Leaders in gas analysis Accurate CO2 infra-red gas analyser Designed for long-term monitoring The EGA30 has a long life sample pump which is designed to give many years of service. The integral soda lime column is used to auto zero the infra-red gas analyser, ensuring long-term stability of gas calibration. The storage capacity of the supplied 2Gb, SD memory card provides enough space for hundreds of thousands of data sets. An integral battery backup delivers enough power for 5 hours of operation should the main electrical supply fail. Data storage can be initiated by a high signal on the CTS line of the RS232, when an analogue input gets to a particular level, after one of the relays actuates or at timed intervals. Information is presented on a large, LED backlit LCD display. Quick and easy installation The EGA30 can be installed and monitoring in minutes. A two part bracket is simply screwed into the wall, all electrical connections can be prewired to a multiway plug, which is easily plugged into a socket on the rear panel. The supplied power connector plugs into a socket on the rear panel and the EGA30 can then be mounted onto the bracket. The whole process can be carried out in 15 minutes. Gas readings, flow indication, power supply status and date and time are all displayed on the main menu screen. Configuration and control is by simple to use, menu driven software

 Open the catalog to page 1
EGA30 Wall Mounted Analyser-2

EGA30 Robust and durable The EGA30’s protective, hinged, front panel gives the membrane overlay protection against the ingress of dirt, dust and water. It also deters unauthorised access to the keypad and memory card. Technical specifications: Pump flow maximum: Flowmeter: Internal Backup battery: Flexible inputs and outputs The EGA30 has seven 12 bit analogue inputs for connection to external sensors or DC voltage signals. These inputs are logged together with the date, time, gas readings, flow, atmospheric pressure and relay status. A single analogue output (voltage or current) can be assigned...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.