ChemSpec_150

ChemSpec_150

ChemSpec_150

Product catalog summary
Overview: The ChemSpec 150 by Bentley Instruments is a fully automated, compact system designed for the determination of Urea (or Milk Urea Nitrogen - MUN) in milk. It utilizes a modified Berthelot reaction to measure urea levels, offering a high throughput of over 150 samples per hour with low cost per sample.
Technical Specifications:
  • Analytical Range: 0-40 mg/dl MUN, 0-86 mg/dl Urea, 0-1.43 mM/dl Urea
  • Speed: 150 samples per hour
  • Sample Intake: 2.5 ml
  • Power: 115 Volts/60Hz/2 Amps or 230 Volts/50Hz/1 Amp
  • Accuracy: Coefficient of Variation (Cv) < 5%
  • Repeatability: Cv < 1.5%
  • Dimensions: Width 63.0 cm, Depth 38.0 cm, Height 27.0 cm, Weight 23.0 kg
Principle of Operation: The ChemSpec 150 operates by dispensing a small amount of milk into a thermostatted well at 40°C, where it is enzymatically split into ammonia and carbon dioxide. After incubation, dye and activator solutions form a green complex, the intensity of which is measured spectrophotometrically. The system uses a 2-channel FIA to differentiate natural and converted ammonia, allowing precise urea concentration determination.
Features:
  • Touchscreen operation for ease of use
  • Automated sample handling with optional barcode readers
  • Data output to screen, printer, or network
  • Integration with standard host-remote solutions
Statistical Test Data:
  • Repeatability Study: Mean, standard deviation, and coefficient of variation for MUN determination in 30 samples
  • Reproducibility Study: Analysis of five milk samples on two instruments over two weeks
  • Accuracy Study: Comparison of MUN concentration determined by enzymatic reference and ChemSpec method
  • Recovery Study: Comparison of target concentration and ChemSpec results for spiked milk samples
Conclusion: The ChemSpec 150 is an accurate, reliable, and efficient instrument for milk urea analysis, suitable for high-throughput laboratory environments.
See more

Catalog excerpts

ChemSpec_150-1

Statistical Test Data: ChemSpec 150 ChemSpec™ 150 1. Repeatability Study MUN [mg/dl] Sd Cu(%) For more information on ChemSpec, please contact Bentley Instruments or your closest representative. Mean, Standard Deviation and Coefficient of Variation for the MUN determination in 30 samples based on five consecutive determinations. BENTLEY INSTRUMENTS Analytical Instruments For The Dairy Industry The ChemSpec 150 utilizes a modified Berthelot reaction to determine the level of Urea in milk. ♦ A small amount of milk is automatically dispensed in a well thermostatted at 40°C. It is then split enzymatically with urease into ammonia and carbon dioxide. ♦ After an incubation time, dye and activator solutions are added to form a green complex. The color intensity is proportional to the ammonia concentration. ♦ The green color intensity is measured spectrophotometrically in a trans-reflectance cell. ♦ The addition of a catalyst in the dye speeds up the reaction, intensifying the green color development, thus enhancing the sensitivity of the method. ♦ The use of a highly precise sample dispenser allows the ChemSpec 150 to bypass a pre-dilution step. The milk is simply diluted in the chemicals used to generate the reactions. This dilution reduces the opacity of the medium, thus improving the spectrophotometric measurement and preventing interference from other milk components. ♦ By alternating measurements of natural ammonia and converted ammonia, the ChemSpec implements a 2-channel FIA. The difference signal from the two measurements allows the ChemSpec to precisely determine the Urea concentration. ♦ A carrousel incubator allows analysis overlap, yielding a capacity of 150 samples per hour.

 Open the catalog to page 1
ChemSpec_150-2

ChemSpec 150 • Accurate • Compact • Fast • Reliable Innovative Design Touch Screen T he ChemSpec 150 touch screen is simple, reliable and accurate. Anyone can operate the instrument after a short orientation. The touch screenis resistant to normal wear and usage, and will not be damaged by wet hands. T ouch screen makes program selection and status checks instantaneous. T he 17-well carrousel allows analysis overlap, resulting in a capacity of 150 samples per hour. A highly precise sample dispenser allows the ChemSpec 150 to bypass a pre-dilution step (milk is diluted in the chemicals used to...

 Open the catalog to page 2

All Bentley Instruments catalogs and technical brochures

  1. DairySpec FT

    8  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.