Website:
bmv-vet
Catalog excerpts
Anesthesia machine installation process 1. Open the package 2. Anesthesia machine instalation (1)、Install wheels and poles Schematic diagram (2)Install the anesthesia machine and tighten the 4 screws. (3) Install evaporation tank a)、put the evaporator on the hanger. Lock switch Lock switch The "lock switch" is in this position and the installation is complete. b)、Rotate the "lock switch" 90° clockwise。 Schematic diagr
Open the catalog to page 12、Pipeline connection steps (1)Anesthesia machine line connection: Exhaust gas discharge line Breathing circuit
Open the catalog to page 24 、Soda lime replacement Turn the screw to replace the soda lime Note: The slot of the soda lime bucket must be aligned, otherwise it will be easy to leak