F-950

F-950

F-950

Product catalog summary
Overview: The F-950 Three Gas Analyzer by Felix Instruments is designed to measure critical ripening gases such as ethylene (C2H4), carbon dioxide (CO2), and oxygen (O2) to maintain optimal produce quality. It is a handheld, lightweight device with a battery life of over 8 hours and Wi-Fi capabilities, making it suitable for various environments.
Key Features:
  • Measures C2H4, CO2, and O2 with results displayed in under 30 seconds.
  • Stores past records on the device and supports multiple languages (English, Spanish, Portuguese).
  • Equipped with a PolarCept!TM water filter to remove non-ethylene hydrocarbons.
  • Display is easily viewable in dimly lit environments.
Applications:
  • Controlled Atmosphere storage facility management.
  • Verification of ethylene mitigation systems.
  • Packing house inspection and precision citrus degreening.
  • Storage assignment in distribution centers and Modified Atmosphere Packaging quality assurance.
Specifications:
  • Sampling flow rate: 70mL/min; Operating environment: 0°C - 50°C.
  • Total sampling time: 80 seconds; Weight: 0.95 kg.
  • Data saving is automated every second; Enclosure made of powder-coated aluminum.
  • Display: Sunlight visible LCD; Power source: Rechargeable Li-ION battery.
  • Dimensions: 18 cm x 13.5 cm x 5.5 cm; Interfaces: USB, WIFI, SD card, Bluetooth.
  • Operating pressure: 500-1200mbar.
Sensor Details:
  • Headspace Sampling:
    • Ethylene: Electrochemical sensor, range 0-200 ppm, resolution 0.1 ppm, response time 35 seconds.
    • CO2: Infrared sensor, range 0-100%, resolution 0.01%, response time 12 seconds.
    • O2: Electrochemical sensor, range 0-100%, resolution 0.10%, response time 12 seconds.
  • Continuous Sampling:
    • Ethylene: Electrochemical sensor, range 0-200 ppm, resolution 0.1 ppm, response time 80 seconds.
    • CO2: Infrared sensor, range 0-100%, resolution 0.01%, response time 6 seconds.
    • O2: Electrochemical sensor, range 0-100%, resolution 0.10%, response time 12 seconds.
Calibration and Maintenance:
  • Offset calibration is required weekly, while span calibration varies from 3 to 12 months depending on the gas.
  • Sensors for ethylene and oxygen are user-replaceable with a lifetime of 1 and 2 years respectively, while the CO2 sensor lasts 5 years.
See more

Catalog excerpts

F-950-1

To achieve ideal conditions in any environment, the F-950 measures critical ripening gases to maintain optimal produce quality at every phase. ► Measures C2H4, CO2 and O2 ► Displays results in under 30 seconds ► Past records stored on the device ► Handheld and lightweight ► Battery lasts for 8+ hours ► Wi-Fi capabilities ► Display easily viewed in dimly lit environments ► PolarCept!TM water filter for removing non-ethylene hydrocarbons ► The F-950 supports English, Spanish, and Portuguese on the instrument display. Applications ► Controlled Atmosphere storage facility management ► Verification of ethylene mitigation systems ► Packing house inspection ► Precision citrus degreening ► Storage assignment in distribution centers ► Modified Atmosphere Packaging quality assu

 Open the catalog to page 1
F-950-2

Sensor Type: Electrochemical Lower Detection Limit: 0.15 ppm Offset Calibration: Weekly User Replaceable Sensor: yes_ Weekly Weekly Sampling flow rate Headspace Sampling Continuous Sampling Sensor Type: Electrochemical Lower Detection Limit: 0.2 ppm Offset Calibration:_Weekly_ User Replaceable Sensor: yes_ www.felixinstruments.com [email protected] Phone: +1 (360) 833-8835 Toll Free: 1-800-767-0119 Fax: +1 (360) 833-1914 Made in USA

 Open the catalog to page 2

All Felix Instruments Inc. catalogs and technical brochures

  1. F-940

    2  Pages

  2. F-900

    2  Pages

  3. F-920

    2  Pages

  4. F-750

    2  Pages