1. Catalogs
  2. lifeassays
  3. Operating manual
video corpo

Operating manual

Operating manual

Operating manual

Product catalog summary
Introduction
The document provides instructions for using the LifeAssays® Canine C-Reactive Protein (CRP) Test Kit with the VetReader instrument. It includes step-by-step procedures for conducting the test and troubleshooting common issues.

Test Procedure
  • Step 1: Insert the disposable algorithm chip into the VetReader and turn it on. The instrument will perform a reagent chip control and stabilize for about 10 minutes.
  • Step 2: Once stabilized, the instrument will display options for the number of tests. Press and hold the enter button to start a measurement.
  • Step 3: Collect a 5µl serum sample using the provided glass capillary, add it to the reagent vial, and close the cap.
  • Step 4: Vortex the reagent vial for 30 seconds as indicated by the VetReader.
  • Step 5: Reinsert the reagent vial into the VetReader within 10 seconds. The instrument will count down for 11 minutes before performing measurements.
  • Step 6: After 11 minutes, the instrument will display the CRP result in mg/L.

Troubleshooting
  • STAB SD: Ensure the instrument has warmed up for 10 minutes with the chip inserted. Contact the distributor if the issue persists.
  • Reagent Error: Verify the instrument was turned on with the correct chip and reagent kit. Check the lot number and run a control test.
  • Chip Error 4: Restart the instrument with the chip inserted. Contact the distributor if the error continues.
  • Chip Error 5: Check the expiration date of the reagent kit. Restart the instrument with the chip inserted. Contact the distributor if the error persists.

Contact Information
For further assistance, contact LifeAssays AB at [email protected] or call +46 46 286 54 00.
See more

Catalog excerpts

Operating manual-1

1. Insert the disposable algorithm chip provided with the LifeAssays® Canine CRP Test Kit and turn on the LifeAssays® VetReader. The instrument performs a reagent chip control and displays, “wait 113” (with a numercial countdown from 113). The instrument will stabilize for about 10 minutes and display during this time, “Stab T”. 2. Once stabilization is complete, the instrument will display, “Canine CRP Tests 20 or 10”. To start a measurement press and hold the enter button. The instrument will display, “wait” followed by “Insert vial”. Load a reagent vial into the VetReader. The instrument will display, “wait 5” before displaying “Collect + Add Sample”. 3. “Collect + Add Sample” will be displayed on the VetReader. Collect a 5pl serum sample using the glass capillary provided with the kit. Unload the reagent vial from the VetReader. Open the cap of the reagent vial. Drop the capillay into the reagent vial and close the cap. 4. “Vortex 30” will be displayed on the VetReader. Vortex the reagent vial for 30 seconds using the timer on display. Once the countdown has reached “0” the instrument will display, “Vortex OK Insert vial”. 5. “Insert Vial 10” will be displayed on the VetReader with a numerical countdown from 10 seconds. Load the reagent vial back into the VetReader before the display shows “0”. The instrument will display “Wait 660” and count down from 660 seconds (11 minutes). The user is now free to leave the instrument! 6. After 11 minutes the instrument will automatically perform 5 measurements, by moving the reagent vial in and out of the internal coil. The result is shown on the display in mg/L and will remain on the display until the reagent vial is unloaded from the instrument. The display on the instrument will show, “Canine CRP XX mg/L” where XX indicates a numerical value. 4 LifeAssays BLOOD ANALYSIS FOR LIFE LifeAssays AB | Solvegatan 43 A, SE-223 70 Lund Sweden | [email protected] | Phone: +46 46 286 54 00 |

 Open the catalog to page 1
Operating manual-2

Trouble Shooting - LifeAssays® Canine C-Reactive Protein (CRP) Reagent error Actions Make sure instrument been warming up for 10 min and chip was inserted before instrument was turned on. If instrument although displays Stab SD contact your distributor for resolution. Reagent/Instrument error. The calculated result is below the set zero for the assay. Make sure instrument was turned on with chip inserted and that chip correspond with the reagent kit it comes with, check Lot number on chip and kit box. Run a control and if CRP value is displayed contact distributor for resolution. Turn off instrument...

 Open the catalog to page 2

All Lifeassays catalogs and technical brochures

  1. Vortex flyer

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.