Capel Capillary electrophoresis system Determination of water-soluble forms of inorganic and organic anions in soils INTRODUCTION The method allows determination of water-soluble forms of inorganic (chloride, sulfate, nitrate, fluoride, and phosphates) and organic (acetate, formate, and oxalate) anions in soils, clays, peats, wastewater silts, activated sludges, and bottom sediments by capillary electrophoresis method. MEASUREMENT METHOD The measurement method is based on the extraction of water-soluble forms of anions by distilled water from a soil sample and subsequent separation, identification, and determination of analyzed components by the capillary electrophoresis method with indirect anion detection at wavelengths of 254 nm. MEASUREMENT RANGE The measurement ranges for the components are presented in the table below. EQUIPMENT AND REAGENTS The Capel capillary electrophoresis system is used in measurements. Data acquisition, collection, processing, and output are performed using a personal computer running under Windows® operating system with Elforun software installed. Lumex Instruments set, order No. 0300001824. EXAMPLES OF REAL ANALYSES BGE: chromate, with DEA and CTA-OH Capillary: Leff/ Ltot 50/60 cm, ID 75 pm Injection: l5o mbar x s Voltage: -25 kV Detection: 254 nm Sample: soil (water extract) Found (mg/kg): The information and specifications in this publication are subject to change without notice.