Installation-/User Manual Sendot Research Optical oxygen sensor System info FluoMini Pro type Firmware Baudrate
Open the catalog to page 1Sendot Research
Open the catalog to page 21. General 1.1. Product Product Versionnummer Software version Firmware version : FluoMini Pro Optical Oxygen Sensor : version 1 : v1.0 : v1.0 170320 Sendot Research FluoMini Pro Fiber probe version (Water or Air) 1.2. Terms of Delivery GENERAL RESEARCH, ADVICE, SALES, DELIVERY AND PAYMENT CONDITIONS OF SENDOT RESEARCH BV (registered under No. 62488295 at KvK Haaglanden) This product has a warranty of two years on the mechanics and electronics (excl. battery). The sensitive coating can be used for a period depending on the measurement frequency and environment where the sensor has to...
Open the catalog to page 41.3. Application This sensor is suitable for measurements in water/fluids and air/gas. The sensor is not suitable for measurements in strongly oxidizing, strongly acidic (pH<1) and strongly caustic (pH>12) media. In general it is advised to test the sensor in the required application. Sendot Research 1.4. Specifications Specifications Measuring range O2 in gas Temperature operation range Pressure operation range Accuracy O2 at > 0% O2 < 1% O2 Accuracy O2 at > 1% O2 < 25% O2 Accuracy O2 at > 25% O2 Accuracy Temperature Accuracy Pressure Resolution O2 Resolution Temperature Resolution...
Open the catalog to page 52. Installation 2.1. Mounting This sensor can be used as a portable and/or table top instrument. Optional items are: • Stand • Mount • Custom made stand or mount For a complete list of optional items please visit http://www.sendot.nl/webshop.html Sendot Research The sensor can be operated through the USB connector on a PC. 2.3. Process connections The FluoMini PRO fiber optic version has a stainless steel probe with a diameter of 6 mm. Stainless steel probe 2.4. Pressure sensor The pressure sensor is positioned inside the sensor-case. An optional outside pressure measurement connection is...
Open the catalog to page 6SentiM research 2.6. Starting screen and buttons The picture below shows the opening screen of the FluoMini Pro O2 sensor and the button layout and numbering applied throughout this manual. System Info fiuoMr* type 300 firmware vl 0 170405 Batt yen 396 Baudrate 19200
Open the catalog to page 73. Measuring with the FluoMini O2|T 3.1. Measurement principle The Optochemical O2|T sensor has two active components embedded in a polymer host. One component measures the temperature of the coating necessary for compensation of the oxygen measurement. The other measures the Oxygen partial pressure. The total pressure of a gas is the sum of the so-called partial pressures of the components in the gas. A standard atmosphere is composed of Oxygen, Nitrogen, some CO2, Water and some trace gasses. (formula 1) Sendot Research ݑ ðݑpݑܰݑpݑΰݑ = ٰݑ ݰݑ2 + °ݑ ݰݑ2 + pݑ ݰݑаݑ2 + ܰݑݢĎ2ݑ + ܢ& .. The O2|T...
Open the catalog to page 83.4.1 Time and date 1. To change the time and date hold button 1 for 2 seconds to enter the main menu. 2. Press enter to enter General Settings 3. Enter date/time by pressing enter 4. By pressing buttons 2, 3 and 4 the desired time and date can be input. By pressing next (button 2), the steps are saved. Power Management To change the power management hold button 1 for 2 seconds to enter the main menu. Enter general settings Enter power management settings By moving the cursor with button 3 or 4 the FluoMini Pro can be kept in sleeping or awake mode. By pressing Set (button 2) the changes...
Open the catalog to page 93.5.1 Data logging Default interval: log every 5 seconds To use the logging facilities the FluoMini Pro has to be setup Go to the main menu and hold button 1 for 2 seconds Press down (button 3) and go to logger settings and press Enter (button 2) Go to Logger On/Off and press Enter (button 2) Change the settings from “off’ to “on” by moving the arrow (button 4) and press Set (button 2) The default setting for the logger is 5 second if no changes are desired hold exit (press 1) for 2 seconds to exit to main screen. (If the interval time has to be changed see 3.5.2) Once you are on the main...
Open the catalog to page 10Go to the main menu and hold button 1 for 2 seconds Press down and go to Logger settings and press enter Press down twice and go to Memory and enter By moving button 4 once select “Yes” and press Set (The sensor will erase the memory) Sensor Settings Sendot Research 3.6.1 Automatic Calibration 1. To calibrate the sensor press button 1 for 2 seconds to enter the main menu 2. Go to sensor settings 3. Select one-point or two-point calibration to setup the temperature and oxygen parameters a. One point calibration press button 3 or 4 to setup the desired temperature setting. The sensor will...
Open the catalog to page 11Go to the main menu by pressing button 2 for 2 seconds. Drop down to sensor settings and enter By pressing button 3 and 4 the desired slope and offset can be set for O2 or Temperature. Slopes are in %. Offset of temperature is degrees multiplied by 10, Offset of oxygen is %O2 multiplied by 100. After desired slope and/or offset has been setup press Next (button 2) to save the changes. filter at. data Set Stop* & Offset Slope Offset Sendot Research 3.6.3 Measuring settings Measuring units The sensor offers several measurement units to display the results 3) %saturation (water) To change the...
Open the catalog to page 124. Operating the FluoMini Pro O2|T via Software Control panel for the sensor Temperature measurements
Open the catalog to page 134.1. Installing the FluoMini Software Sendot Research To install the FluoMini Pro package please follow the following steps in the installation wizard.
Open the catalog to page 14Sendot Research
Open the catalog to page 15Measuring with the FluoMini Pro Note: Please keep in mind that the software operates as an extension of the sensor. Meaning the sensor should operate either via the software or via the display on the housing. When the sensor is connected to a PC it is advised to avoid operation with the sensor itself. Before measuring Sendot Research 1. Plug in the sensor to a desired USB port. 2. Wake up the device by pressing any of the buttons on the housing. 3. Start the FluoMini Pro software. 4.2.1 Single Measurement 1. Press Measure Once to perform a single measurement. 2. Press Save Measurement to...