FluoMini - Oxygen Sensor
11Pages

{{requestButtons}}

Catalog excerpts

FluoMini - Oxygen Sensor - 1

Installation-/User Manual FluoMini O2IT optical oxygen sensor

Open the catalog to page 1
FluoMini - Oxygen Sensor - 3

1. General 1.1. Product Product: FluoMini Optical Oxygen Sensor Versionnummer: version 2 Software version: 8.5 Fiber probe version (Water or Gas) Non Invasive Oxygen Version 1.2 Leveringsvoorwaarden The "ALGEMENE ONDERZOEKS-, ADVIES-, VERKOOP-, LEVERINGS- EN BETALINGSVOORWAARDEN VAN SENDOT RESEARCH BV (gedeponeerd onder nr. 62488295 bij KvK Haaglanden)" are applicable tot his product.

Open the catalog to page 3
FluoMini - Oxygen Sensor - 4

Semin research 1.3 Warranty This product has a warranty of two years on the mechanics and electronics. The sensitive coating can be used for a period depending on the measurement frequency and environment where the sensor has to operate in, like temperature, pressure and average oxygen partial pressure. 1.4 Toepassing This sensor is suitable for measurments in water/fluids and air/gas. The sensor is not suitable for measurements in strongly oxidizing, strongly acidic (pH<1) and strongly caustic (pH>12). In general it is advised to test the sensor in the required application. 2....

Open the catalog to page 4
FluoMini - Oxygen Sensor - 5

Supply Voltage 3. Installation 3.1 Montage This sensor is suitable for table top usage, an optional stand is available. It is also possible to mount the sensor on a stand. Special mounts are available on request. 3.2 Connections The sensor can be operated though the USB connector on a PC. A USB “on the go” cable makes it possible to operate the sensor from a tablet (Android/Windows). De sensor kan vanuit een Usb aansluiting op een PC aangesloten worden. Via een USB “on the go” kabel kan de sensor ook vanuit een tablet gebruikt worden (Android/Windows) 3.3 Proces connections De fiber optic...

Open the catalog to page 5
FluoMini - Oxygen Sensor - 6

4. Measuring with the FluoMini O2|T 4.1 Measurement values The Optochemical O2|T sensor has two active components embedde in a polymer host. One component measure the temperature of the coating necessary for compensation of the oxygen measurment. The other measure the Oxygen partial pressure. The total pressure of a gas is the sum of the so-called partial pressures of the components in the gas (formula 1) ݑ ðݑpݑܰݑpݑΰݑ = ٰݑ ݰݑ2 + °ݑ ݰݑ2 + pݑ ݰݑаݑ2 + ܰݑݢĎ2ݑ + ܢ& .. Because the O2|T coating can be considered as a small gas compartment, in gas as well as in water. Basically it makes no...

Open the catalog to page 6
FluoMini - Oxygen Sensor - 7

The box top-left is a status box giving extra information what is going on. The button <Start> starts the measurements. The status of this toggle button will change from <start> to <stop> if clicked and from <stop> to <start> if clicked once again. The results from the measurements will be stored in a file in the C:\Data directory. The name of this file can be adapted by changing the text in the <extra file identifier> field. The program should first be stopped to make the change effective. With the <Graph> button, a graph of the measured data will be presented. Below the graph a choice is...

Open the catalog to page 7
FluoMini - Oxygen Sensor - 8

The screen shots bellows show the successive tabs with a short explanation The various oxygen measurement units

Open the catalog to page 8
FluoMini - Oxygen Sensor - 9

Choice in storing each or each 10th or 50th datapoint. Time interval setting (secondes) Raw data stored checkbox (default) Choose a specific temperature Change the slope and offset of the calibration of the sensor for both temperature and oxygen. The offset to be entered in the textbox and should calibrated. These can be stored inside the sensor.

Open the catalog to page 9
FluoMini - Oxygen Sensor - 10

Automatic calibration routine for zero oxygen and 100% air (see maintenance) 5. Onderhoud/Kalibratie This sensor does not need any maintenance besides regular check-ups and calibration if necessary. If the coating does not fulfil its specifications it should be replaced and/or if necessary recalibrated. Calibration Procedure: Place the sensor probe in a zero oxygen medium (100% Nitrogen gas flow) and leave it to stabilize for several minutes. If a DO sensor has to be calibrated, lead the 100% Nitrogen through a moisturized medium before it reaches the sensor probe). Click the button several...

Open the catalog to page 10
FluoMini - Oxygen Sensor - 11

Adres Sendot research BV Veilingweg 44 3981 PC Bunnik Nederland Telefoon +31 (0)30-636 84 77 E-mail info@sendot.nl